18msMIN (at 10kHz) 20kHz~5.00MHz 40ms max 42Hz~19.0kHz 40ms ~
Measurement circuit mode
Series equivalent or Parallel equivalent.
Zero correction
You can automatically correct parasitic impedance with the circuit open or short-circuited for zero balancing. Zero correction in open mode is executable with 1k or higherthat in short mode, with less than 1k
Memory
9 testing conditions storable in memory
GO/NO Judgment
High/low limits can be set in a range of 00000 to ±19999each subject device can be classified into 10 sorts according to the judgment (absolute value or percentage).
Display
DISP-A/DISP-B 4-1/2digit display
Judgment results output
DISP-A/DISP-B LO, HI, GO, TOTAL GO, TOTAL NO-GO; BUZZER ON (Isolated)
Output signals
MEASURE END, TEST BUSY, BIN, ERROR (Isolated)
Guaranteed temperature range
23°C ±5°C (Specified accuracies are guaranteed in this temperature range. In a range of 5°C to 40°C, limits of accuracy are doubled.)
External control signals
CONTROL PANEL/EXT, EXT TRIGGER, PANEL No (Isolated)
Signals
RS-232C standard provision
Operating temperature range
0~50°C
Power requirements
100/120/220/240V AC ±10%, 50/60Hz; approx. 60VA
Dimensions & Mass
250(W) x 148(H) x 400(D) mm, approx. 7kg
Options