Applications
Measurement and test benches
Calibration technology
Laboratories
Plant construction and machine building
Special features
Accuracy 0.1 %, without additional temperature error in a range of 10 ... 60 °C
Optional accuracy of 0.05 % (full scale) available
Fast measuring rates up to 1 kHz
Analogue, USB and CANopen? output signals available
On-site calibration possible using product software
Description
Precise
The model P-30 and P-31 pressure transmitters have been developed for precision measurements. Through the use of special WIKA pressure sensors, precision measurements with a maximum measuring deviation of as low as 0.05 % of span are guaranteed. As a result of their active temperature compensation, these pressure transmitters have no additional temperature error in the range of 10 ... 60 ° C.
Fast
The high measuring and output rates of up to 1 kHz make the measured value available as quickly as possible.
Compact
The compact design makes the pressure transmitter ideal for mounting into test benches, such as 19" racks.
Versatile
The models P-30 and P-31 offer a wide selection of electrical connections, process connections and measuring ranges, as well as a large number of different output signals. In addition to the standard analogue signals, USB and CANopen? versions are also available.
Via a USB service interface and the WIKA configuration software "EasyCom", the models P-30 and P-31 can quickly and easily be adjusted on site.
Thanks to the simple-to-use software "Wika data logger“, the USB version can also be used to save measured values and create customised reports.
Measuring ranges
The given measuring ranges are also available in mbar, kg/cm2 and MPa.
Other measuring ranges on request
Vacuum resistance
Yes
Output signal
Voltage supply
Power supply
The permissible power supply depends on the corresponding
output signal.
Total current consumption
The total current consumption is dependent on the respective signal type.
Load
Accuracy
Accuracy at reference conditions
Standard: ≤ ±0.1 % of span
Option: ≤ ±0.05 % of span 1)
1) For +/- measuring ranges and measuring range ≤ 0.4 bar on request
Including non-linearity, hysteresis, non-repeatability, zero offset and end value deviation (corresponds to measured error per IEC 61298-2). Calibrated in vertical mounting position with process connection facing downwards.
Non-linearity (per IEC 61298-2)
≤ ±0.04 % of span BFSL
Temperature error
In the range of -20 ... +80 °C the instrument is actively compensated.
Total error band (10 ... 60 °C)
≤ ±0.1 % of span
Long-term stability
≤ ±0.1 % of span/year
Adjustability
Adjustment via the "EasyCom 2011" or "EasyCom CANopen® software
Zero point: -5 ... +10 % of span
Span: -50 ... +5 % of span
Measuring rate
The measuring rate is dependent on the respective signal type.
Reference conditions
Temperature
15 ... 25 °C
Atmospheric pressure
860 ... 1,060 mbar
Humidity
45 ... 75 % relative
Power supply
DC 24 V
DC 5 V with USB version
Warm-up time
< 10 min
Mounting position
Process connection lower mount (LM)
Operating conditions
Ingress protection (per IEC 60529)
The ingress protection depends on the type of electrical connection.
The stated ingress protection only applies when plugged in using mating connectors that have the appropriate ingress protection.
Vibration resistance
10 g (IEC 60068-2-6, under resonance)
Shock resistance
200 g (IEC 60068-2-27, mechanical)
Service life
10 million load cycles
Free fall test
The instrument is resistant to an impact onto concrete from a height of 1 m.
Temperatures
|